Document information
- Document type
- Non-binding MDCG guidance
- Issuing body
- Medical Device Coordination Group (MDCG)
- Issued
- November 2026
- Official source
- Read the official source
Which devices the guidance addresses
MDCG 2026-6 focuses on orphan IVDs and indications meeting its criteria. Manufacturers should explain intended purpose, target population, indication and available alternatives. Limited sample availability alone does not establish orphan status.
How to address limits in pre-market data
Identify the performance claims and gaps, then explain what the available clinical and non-clinical evidence can support. If extrapolating from other populations or indications, document comparability, limitations and residual uncertainty in the performance evaluation report.
How to plan post-market evidence
Use a device-specific post-market performance follow-up plan with questions, data sources and review arrangements. Feed the findings back into performance evaluation, risk management and product information.
What each party does
The manufacturer prepares the applicability rationale and performance evidence. The notified body assesses it under the applicable procedure. Expert panel advice may be relevant in the situations described in the guidance. The guidance is non-binding and conformity assessment remains governed by the IVDR.